Extracorporeal ultrasound exposure by the low-frequency acoustic amplitude-modulated signal on a ureteral stent for preventing its incrustation: experimental determination of optimal application points

https://doi.org/10.21886/2308-6424-2021-9-3-70-78

Abstract

Introduction. Incrustation and biofilms formation on the surface of ureteral stents are still the most significant complications of internal drainage of the upper urinary tract. There are much researchers conducted to combat these complications. The lack of a solution to this problem affects the ultimate results of treatment and economic losses. The issue of impact by physical methods on the ureteral stent, particularly the use of extracorporeal ultrasound acoustic exposure remains, promising and poorly covered.

Purpose of the study. To determine the optimal application points of extracorporeal acoustic exposure by the low-frequency ultrasonic amplitude-modulated signal on a ureteral stent in an experiment.

Materials and methods. The original device was designed. The main principle of its operation is the generation of an amplitude-modulated ultrasonic signal in two modes: pulsed and permanent. A sexually mature mongrel dog was an experimental animal. The ureteral stent was placed by laparotomy and cystotomy. Intraoperatively, the emitter of the developed device was applied to the skin of the animal, according to the previously indicated topographic and anatomical landmarks. At the same time, an ultrasonic wave noise analyzer was applied through the laparotomy wound to the appropriate level of the ureter. Measurements of ultrasound intensity indicators were performed three times in two operating modes of the device.

Results. Pulsed mode: for the ureteral upper third, the highest ultrasonic intensity (123.67 dB) was achieved along the posterior axillary line. For the ureteral middle third, the best ultrasound intensity (115 dB) was obtained by the posterior axillary line. For the ureteral lower third, the highest ultrasound intensity (113.67 dB) was noted along the middle axillary line.

Permanent mode: the best ultrasonic intensity in the projection of the ureteral upper, middle, and lower thirds was achieved along the posterior axillary line and was 118.67 dB, 117 dB and 116.67 dB, accordingly. However, there was an excessive heat effect, manifested by hyperemia and hyperthermia of the animal's skin, fascicular muscle contractions during the instrument functioned in the permanent mode, which can potentially lead to thermal burns and intolerance to the procedure.

Conclusion. The pulsed mode of the device function is most safe. The optimal application points of the instrument emitter for the ureteral upper and middle thirds is the posterior axillary line, and for the ureteral lower third is the middle axillary line.

About the Authors

A. Yu. Tsukanov
Omsk State Medical University
Russian Federation

Anton Yu. Tsukanov – M.D., Dr.Sc.(M), Full Prof.; Head, Dept. of Surgery and Urology

644099, Omsk, 12 Lenin st.

tel.: + 7 (913) 967-36-76


Competing Interests:

The authors declare no conflicts of interest.



D. S. Akhmetov
Omsk State Medical University
Russian Federation

Daniyar S. Akhmetov – M.D.; Postgraduate Student, Head, Dept. of Surgery and Urology

644099, Omsk, 12 Lenin st.


Competing Interests:

The authors declare no conflicts of interest.



A. A. Novikov
Omsk State Technical University
Russian Federation

Aleksei A. Novikov – Dr.Sc.(Tech.); Prof., Dept. of Mechanical Engineering and Materials Science

644050, Omsk, 11 Mira ave.


Competing Interests:

The authors declare no conflicts of interest.



N. A. Negrov
Omsk State Technical University
Russian Federation

Dmitry A. Negrov – Cand.Sc.(Tech.); Assist.Prof., Dept. of Mechanical Engineering and Materials Science

644050, Omsk, 11 Mira ave.


Competing Interests:

The authors declare no conflicts of interest.



A. R. Putintseva
Omsk State Technical University
Russian Federation

Aleksandra R. Putinceva – Postgraduate Student, Dept. of Mechanical Engineering and Materials Science

644050, Omsk, 11 Mira ave.


Competing Interests:

The authors declare no conflicts of interest.



S. P. Semikina
Omsk State Medical University
Russian Federation

Sofiia P. Semikina – Resident, Dept. of Surgery and Urology

644099, Omsk, 12 Lenin st.


Competing Interests:

The authors declare no conflicts of interest.



References

1. Ordonez M, Hwang EC, Borofsky M, Bakker CJ, Gandhi S, Dahm P. Ureteral stent versus no ureteral stent for ureteroscopy in the management of renal and ureteral calculi. Cochrane Database Syst Rev. 2019;2(2):CD012703. DOI: 10.1002/14651858.CD012703.pub2

2. Trusov P.V., Gusev A.A. Treatment of kidney stones: standards and innovations. Vestnik Urologii. 2019;7(2):93-111. (In Russ.). DOI: 10.21886/2308-6424-2019-7-2-93-111

3. Zumstein V, Betschart P, Albrich WC, Buhmann MT, Ren Q, Schmid HP, Abt D. Biofilm formation on ureteral stents – Incidence, clinical impact, and prevention. Swiss Med Wkly. 2017;147:w14408. DOI: 10.4414/smw.2017.14408

4. Pengfei S, Yutao L, Jie Y, Wuran W, Yi D, Hao Z, Jia W. The results of ureteral stenting after ureteroscopic lithotripsy for ureteral calculi: a systematic review and meta-analysis. J Urol. 2011;186(5):1904-9. DOI: 10.1016/j.juro.2011.06.066

5. El-Azizi M, Rao S, Kanchanapoom T, Khardori N. In vitro activity of vancomycin, quinupristin/dalfopristin, and linezolid against intact and disrupted biofilms of staphylococci. Ann Clin Microbiol Antimicrob. 2005;4:2. DOI: 10.1186/1476-0711-4-2

6. Liatsikos E, Kallidonis P, Stolzenburg JU, Karnabatidis D. Ureteral stents: past, present and future. Expert Rev Med Devices. 2009;6(3):313-24. DOI: 10.1586/erd.09.5

7. Kogan M.I., Moisyuk Ja. G., Shkodkin S.V., Saidulaev D.A., Idashkin Yu. B. Effectiveness of ureteral stents with nanostructured coating in renal transplantation (preliminary results). Urologiia. 2015;1:58-61. (In Russ.). eLIBRARY ID: 23608466

8. Cauda F, Cauda V, Fiori C, Onida B, Garrone E. Heparin coating on ureteral Double J stents prevents encrustations: an in vivo case study. J Endourol. 2008;22(3):465-72. DOI: 10.1089/end.2007.0218

9. Cadieux PA, Chew BH, Knudsen BE, Dejong K, Rowe E, Reid G, Denstedt JD. Triclosan loaded ureteral stents decrease proteus mirabilis 296 infection in a rabbit urinary tract infection model. J Urol. 2006;175(6):2331-5. DOI: 10.1016/S0022-5347(06)00252-7

10. Gultekinoglu M, Kurum B, Karahan S, Kart D, Sagiroglu M, Ertaş N, Haluk Ozen A, Ulubayram K. Polyethyleneimine brushes effectively inhibit encrustation on polyurethane ureteral stents both in dynamic bioreactor and in vivo. Mater Sci Eng C Mater Biol Appl. 2017;71:1166-74. DOI: 10.1016/j.msec.2016.11.125

11. John T, Rajpurkar A, Smith G, Fairfax M, Triest J. Antibiotic pretreatment of hydrogel ureteral stent. J Endourol. 2007;21(10):1211-6. DOI: 10.1089/end.2007.9904

12. Wang L, Yang G, Xie H, Chen F. Prospects for the research and application of biodegradable ureteral stents: from bench to bedside. J Biomater Sci Polym Ed. 2018;29(14):1657-66. DOI: 10.1080/09205063.2018.1498184

13. Tunney MM, Jones DS, Gorman SP. Biofilm and biofilmrelated encrustation of urinary tract devices. Methods Enzymol. 1999;310:558-66. DOI: 10.1016/s0076-6879(99)10043-0

14. Hazan Z, Zumeris J, Jacob H, Raskin H, Kratysh G, Vishnia M, Dror N, Barliya T, Mandel M, Lavie G. Effective prevention of microbial biofilm formation on medical devices by low-energy surface acoustic waves. Antimicrob Agents Chemother. 2006;50(12):4144-52. DOI: 10.1128/AAC.00418-06

15. Kopel M, Degtyar E, Banin E. Surface acoustic waves increase the susceptibility of Pseudomonas aeruginosa biofilms to antibiotic treatment. Biofouling. 2011;27(7):701-10. DOI: 10.1080/08927014.2011.597051

16. Dror N, Mandel M, Hazan Z, Lavie G. Advances in microbial biofilm prevention on indwelling medical devices with emphasis on usage of acoustic energy. Sensors (Basel). 2009;9(4):2538-54. DOI: 10.3390/s90402538

17. Patent RF na izobretenie № 2703472/17.10.2019. Bjul. №29. Novikov A. A., Tsukanov A. Yu., Akhmetov D. S. Method of non-invasive sanitation of ureteral stents (In Russ.). Available at: https://edrid.ru/en/rid/219.017.d836.html Accessed August 20, 2021


Review

For citations:


Tsukanov A.Yu., Akhmetov D.S., Novikov A.A., Negrov N.A., Putintseva A.R., Semikina S.P. Extracorporeal ultrasound exposure by the low-frequency acoustic amplitude-modulated signal on a ureteral stent for preventing its incrustation: experimental determination of optimal application points. Urology Herald. 2021;9(3):70-78. (In Russ.) https://doi.org/10.21886/2308-6424-2021-9-3-70-78

Views: 1013


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2308-6424 (Online)